FPP Ceramic Plate Temperature Control Unit
- −5 … +80 °C setting range
- PID control, ±0.1 °C precision
- 4 × 4 cm ceramic heating/cooling plate

Four-point conductivity measurement system combining the probe, measurement electronics and −5…+80 °C temperature control in one portable unit
Manufacturer: Entek Elektronik, İvedik OSB, Ankara · Designed and manufactured in TürkiyeThe FPP 510 combines the four point probe head, microprocessor-controlled measurement electronics and a ceramic-plate temperature control unit in a single portable instrument. The probe arm on a vertical slide is positioned with a screw mechanism onto samples of different thickness and size (thin film, thick plate, polymer, coating, textile); the PTFE probe head with 99.9% platinum pins (pin spacing 1 mm, pin diameter 0.5 mm) on a 4 × 4 cm hard ceramic base minimizes contamination and is easy to clean.
Depending on the selected mode, results are read directly on the LCD as Ω/□ (sheet resistance), Ω·cm (bulk/volume resistivity) or Ω (resistance), each with the simultaneous conductivity value in S/cm. Sample thickness (d) and the geometry-dependent correction factor (k) are entered from the keypad; the instrument displays the value computed according to the correction tables in the literature (Topsøe, 1966). 16-bit sampling resolution and averaging of up to 99 samples give stable results; the 20 nA – 10 mA current range protects sensitive, low-conductivity samples, and automatic ranging finds the highest resolution.
The integrated PID temperature control unit (±0.1 °C precision, digital display) holds the ceramic plate between −5 and +80 °C, so the temperature dependence of conductivity (polymers, ionic conductors, semiconductors) is measured on the same setup. 10 Ω / 1 kΩ / 10 kΩ / 1 MΩ reference resistors are supplied with the instrument; in TEST mode, measurement accuracy is checked instantly against them.
| Probe head | PTFE; 4 × platinum (99.9%) pins; spacing 1 mm; pin diameter 0.5 mm |
|---|---|
| Sample base | 4 × 4 cm hard ceramic plate (heating/cooling surface) |
| Mechanics | Screw-driven probe arm on a vertical slide; different thicknesses and sizes |
| Measurement modes | Sheet resistance (Ω/□), bulk/volume resistivity (Ω·cm), resistance (Ω); simultaneous S/cm |
| Current range | 20 nA – 10 mA; automatic/manual ranging |
| Resistance range | 0.1 Ω – 200 MΩ |
| Resolution | 16-bit sampling; averaging of up to 99 samples |
| Parameters | Sample thickness d (mm), correction factor k |
| Temperature control | −5 … +80 °C (depending on ambient and sample); PID, ±0.1 °C; digital display; ON/OFF and SET keys |
| Thickness limit for sheet resistance | ≤ 0.6 mm (t ≤ 0.4·s); bulk resistivity for thicker samples |
| Display | LCD (result + mode) and integrated keypad |
| Calibration | Supplied 10 Ω / 1 kΩ / 10 kΩ / 1 MΩ reference resistors (TEST mode) |
| Power | Mains; rear-panel on/off switch |
Source: Entek Elektronik product documents and user manuals. Specifications are subject to change without notice.
Accessories manufactured or supplied by Entek for use with this instrument; the specifications of each are on its own page.




Our technical team evaluates the configuration and alternatives according to your measurement needs.