Curriculum · Module 2 · Beginner–Intermediate

Voltammetric and Chrono-techniques

From cyclic voltammetry to pulsed techniques, from chronoamperometry to rotating electrodes: the everyday tools of electrochemistry.

What is cyclic voltammetry (CV)? How is a voltammogram read?

Frequently searched:what is cyclic voltammetryinterpreting a CV voltammogramRandles-Sevcikeffect of scan rate

1 Theory

Excitation and response

In CV the potential is scanned as a triangular wave between two limits at constant rate (v, V/s) and the current is recorded. For a reversible couple, during the forward scan the species at the surface is depleted while the diffusion layer grows; the current first rises, then decreases under transport limitation — this is how the peak forms. On the reverse scan the product is converted back and a peak in the opposite direction is seen.

Diagnostic criteria

In a reversible (Nernstian) system: ΔE_p = E_pa − E_pc ≈ 59/n mV (25 °C), peak-current ratio i_pa/i_pc ≈ 1, peak potentials independent of scan rate and peak current proportional to v^(1/2) according to the Randles–Ševčík equation: i_p = 2.69×10⁵·n^(3/2)·A·D^(1/2)·c·v^(1/2). In a quasi-reversible system ΔE_p grows with scan rate (k⁰ is calculated by the Nicholson method); in an irreversible system the reverse peak disappears. For surface-attached (adsorbed) species the peak current is directly proportional to v and ideally ΔE_p = 0 — this criterion separates a diffusion-controlled process from a surface-controlled one.

Mechanism diagnosis

Coupled chemical reactions (EC, ECE, catalytic EC′) change the peak ratios and potential shifts in characteristic ways; a scan-rate series is the basic tool for distinguishing these mechanisms. Overlaying experimental curves with digital simulation (finite differences) allows quantitative determination of k⁰, α and chemical rate constants.

Practical settings

Scan rates range from 1 mV/s (solid state, batteries) to 100 V/s+ (microelectrodes, fast kinetics); at high rates the capacitive current (i_C ∝ v) swamps the Faradaic current (∝ v^(1/2)) and the signal-to-background ratio falls. Potential step (resolution), filters and ohmic-drop compensation directly affect the peak shape. On battery electrodes CV is a fast map of redox potentials and phase transformations; on supercapacitors a near-rectangular curve indicates ideal capacitive behaviour.

2 Application in measurement

In CV the current is recorded while the potential is scanned back and forth at constant rate. For a reversible redox couple the peak separation is about 57/n mV and the peak current is proportional to the square root of the scan rate (Randles-Ševčík); for adsorbed species the peak current is directly proportional to the rate. Peak potentials, reversibility, electron number and accompanying chemical steps (EC, ECE) are extracted from the voltammogram.

Practical rules: record the potential window and the blank voltammogram of the supporting electrolyte first; choose the scan rate so that it does not increase ohmic drop; measure at different rates to separate capacitive from Faradaic current. An internal standard such as ferrocene corrects reference drift in non-aqueous media.

The EC-Lab® CV Sim/CV Fit tools let you test a mechanism hypothesis by simulation; the Entek potentiostat, with its built-in sweep generator, implements CV in teaching laboratories with a recorder/PC.

Key equations

  • Randles-Ševčík (25 °C): i_p = 2.69×10⁵·n^(3/2)·A·D^(1/2)·C·v^(1/2)

3 Applicability

Where it applies

  • The first 'fingerprint' of a new system: redox potentials, reversibility, electrolyte window.
  • Catalyst ECSA determination, electropolymerisation, phase transformation and SEI formation on battery electrodes.
  • Kinetic parameter determination (scan-rate series + simulation).

Limits and caveats

  • Low sensitivity for trace-concentration analysis; pulsed techniques are preferred.
  • In systems with high ohmic drop the kinetic interpretation can be misleading; R_u must be corrected first.

4 What did the application notes find?

  • Kinetics AN41-1 — The effect of scan rate on peak potential and current in a simple E mechanism is simulated with CV Sim; the transition from reversible to irreversible as k⁰ decreases is followed numerically.
  • Kinetics AN41-2 — The effect of ohmic drop and double-layer capacitance on the CV shape is shown by simulation: R_u enlarges the peak separation, C_dl shifts the curve with a tilted background; together they can mimic slow kinetics.
  • Battery AN26 — Shows how multi-cycle CV and battery data are plotted with customisation in the software and how derived quantities (e.g. capacity, dQ/dV) are generated with the calculator.

5 Which instruments perform this measurement?

References

  1. article N. Elgrishi, K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart, J. L. Dempsey. A Practical Beginner's Guide to Cyclic Voltammetry. J. Chem. Educ. 95, 197–206, 2018. doi:10.1021/acs.jchemed.7b00361 open access
  2. book A. J. Bard, L. R. Faulkner, H. S. White. Electrochemical Methods: Fundamentals and Applications (3rd ed.). Wiley, 2022.
  3. web Bio-Logic Science Instruments. Learning Center — articles, application notes and tutorial videos. biologic.net, 2026. link

When are chronoamperometry (CA) and chronopotentiometry (CP) used?

Frequently searched:chronoamperometrychronopotentiometrypotential stepconstant-current experiment

1 Theory

Chronoamperometry (CA)

The potential is stepped suddenly from one value to another and the time-dependent decay of the current is recorded. In a diffusion-controlled step the current obeys the Cottrell equation: i·√t is constant. The linearity of the i vs t^(−1/2) plot is the measure of diffusion control, its slope the measure of the product n·A·√D·c. The exponential decay in the first microseconds–milliseconds is the charging of the double layer (τ = R_u·C_dl); this region is excluded. In nucleation and growth studies (electroplating, metal deposition) the shape of the current–time curve is compared with instantaneous/progressive nucleation models.

Chronopotentiometry (CP)

A constant current is applied and the potential is followed against time. When the species at the surface is depleted the potential jumps to the next reaction; this transition time τ depends on concentration through the Sand equation: i·τ^(1/2) = nFAc√(πD)/2. In the battery world the counterpart of CP is galvanostatic charge/discharge; potential plateaus correspond to phase transformations, sloping regions to single-phase solid solutions.

Pulse and relaxation measurements

When short current pulses are combined with subsequent rest (relaxation), the ohmic, charge-transfer and diffusion contributions separate by time scale: the first instantaneous jump is R_u, the millisecond–second region charge transfer, longer times diffusion. In batteries HPPC and GITT rest on this principle; the result is the time-domain counterpart of the frequency-domain behaviour measured by EIS (they are linked by the Laplace transform).

2 Application in measurement

In CA the potential is applied as a step and the current is followed in time: in a diffusion-controlled system Cottrell behaviour (i ∝ t^-1/2) is observed; it is the basic technique for electroplating, nucleation-growth and coulometry. In CP a constant current is applied and the potential is followed; battery charge/discharge (GCD) is a special case of this, and the Sand equation defines the transition time.

Initial characterisation is done with CV; once the mechanism and window are established, quantitative parameters (D, charge, efficiency) are obtained with CA/CP. Coulometry (total charge) is used in electrolysis-efficiency and coating-thickness calculations; at high currents the Entek mAs Meter is designed for this purpose.

With conditional limits and loops, CA/CP can be combined with CV and EIS inside an experiment sequence.

Key equations

  • Cottrell: i = nFAC√(D/πt)
  • Sand: τ^(1/2) = (nFAC√(πD))/(2I)

3 Applicability

Where it applies

  • Diffusion-coefficient and concentration determination, nucleation/growth kinetics, calibration of sensors.
  • Internal resistance and pulse power (HPPC) measurement on battery cells, GITT/PITT.

Limits and caveats

  • In long-duration CA convection and edge effects create deviations from Cottrell; at microelectrodes a steady-state current forms and Cottrell loses validity.

4 What did the application notes find?

  • Battery AN38 — Shows that the time-domain 'dynamic resistance' R_Δt measured by current pulse/interrupt corresponds to different frequency regions of EIS depending on the chosen time interval, and that AC and DC results agree within this framework.
  • Battery AN28 — Shows that an insufficient sampling rate in the current-interrupt method leads to an underestimated ohmic resistance, and that with fast data acquisition the result approaches that of EIS.

5 Which instruments perform this measurement?

References

  1. book A. J. Bard, L. R. Faulkner, H. S. White. Electrochemical Methods: Fundamentals and Applications (3rd ed.). Wiley, 2022.
  2. article N. Elgrishi, K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart, J. L. Dempsey. A Practical Beginner's Guide to Cyclic Voltammetry. J. Chem. Educ. 95, 197–206, 2018. doi:10.1021/acs.jchemed.7b00361 open access

Pulsed techniques (DPV, SWV, NPV) and stripping voltammetry: how is trace analysis done?

Frequently searched:differential pulse voltammetrysquare wave voltammetryanodic stripping voltammetrywhat is polarography

1 Theory

Why pulses?

In a linear sweep the capacitive current flows continuously together with the Faradaic current and keeps the detection limit around 10⁻⁵ M. After a potential step the capacitive current decays exponentially (τ = R_u·C_dl, of the order of ms) while the Faradaic current decreases more slowly as t^(−1/2). If the current is sampled at the end of the pulse the capacitive component is largely eliminated; the 10⁻⁷–10⁻⁹ M sensitivity of pulsed techniques comes from this.

Techniques

Normal pulse voltammetry (NPV) applies pulses of increasing amplitude from a base potential; it gives a sigmoid wave. Differential pulse voltammetry (DPV) superimposes constant-amplitude pulses on a slow ramp and plots the difference between the currents before and after the pulse; the result is a derivative-like peak centred on E₁/₂ with a width of 90.4/n mV. Square-wave voltammetry (SWV) applies a forward–reverse pulse pair at each staircase step and takes the difference of the forward and reverse currents; in reversible systems it gives the highest sensitivity and the fastest scan (seconds). The peak current is linear with concentration; peak height and width carry information about reversibility and n.

Stripping voltammetry

In trace-metal analysis the analyte is first deposited on the electrode by reduction during a long preconcentration step (minutes), then 'stripped' by a scan in the positive direction (anodic stripping, ASV). The preconcentration factor gains 10²–10³-fold sensitivity; ng/L levels are attainable. Mercury-film, bismuth-film and boron-doped diamond electrodes are common; adsorptive stripping (AdSV) extends the method to organic and complexing species.

Polarography and the dropping mercury electrode

Heyrovský's polarography is voltammetry performed with the dropping mercury electrode (DME); each drop provides a fresh, reproducible surface and the hydrogen overpotential allows very negative potentials to be reached. Static/hanging mercury drop electrodes (SMDE/HMDE) combined with modern pulsed techniques are still the reference method in heavy-metal analysis; the use of mercury is subject to laboratory safety rules.

2 Application in measurement

Pulsed techniques sample the current at the end of the pulse, largely eliminating the capacitive current and lowering the detection limit to the nM–pM level. DPV gives a peak-shaped response, SWV a fast and highly sensitive one; NPV produces a step-shaped voltammogram in a convection-free medium.

In stripping voltammetry the analyte is first deposited on the electrode (preconcentration) and then measured as it is stripped; it is standard in heavy-metal determination (Pb, Cd, Zn, Cu). Mercury drop electrodes (DME/SMDE/HMDE) and polarography are the classical ground of these techniques; the Entek Polarographic Analyser and CGME are a domestic solution for these applications.

On Bio-Logic instruments DPV/SWV/NPV are available in EC-Lab®; they are used in sensor development and electroanalysis.

3 Applicability

Where it applies

  • Trace metal and organic analyte determination (water, food, pharmaceuticals), electrochemical sensor and biosensor calibration, DNA/protein interaction studies.

Limits and caveats

  • Pulsed techniques offer no advantage in high-current/low-impedance systems such as batteries and corrosion.
  • In irreversible systems the SWV peak height drops; DPV or NPV may be more suitable.

4 What did the application notes find?

  • Sensor AN67 — Compares SWV, DPV and NPV on the same model system; explains the effect of pulse amplitude, pulse duration and frequency parameters on peak height and resolution and the approach to parameter selection for the 10⁻⁶–10⁻⁹ M level.

5 Which instruments perform this measurement?

References

  1. article J. G. Osteryoung, R. A. Osteryoung. Square Wave Voltammetry. Anal. Chem. 57, 101A–110A, 1985. doi:10.1021/ac00279a789
  2. book A. J. Bard, L. R. Faulkner, H. S. White. Electrochemical Methods: Fundamentals and Applications (3rd ed.). Wiley, 2022.
  3. article N. Elgrishi, K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart, J. L. Dempsey. A Practical Beginner's Guide to Cyclic Voltammetry. J. Chem. Educ. 95, 197–206, 2018. doi:10.1021/acs.jchemed.7b00361 open access

Rotating disk (RDE) and ring-disk (RRDE) electrode experiments

Frequently searched:RDE experimentRRDE collection efficiencyKoutecký-LevichORR catalyst measurement

1 Theory

Hydrodynamic principle

The rotating disk draws solution in at the centre and throws it out radially; a time-independent diffusion layer whose thickness is set by the rotation rate forms at the disk surface. This is one of the rare experimental geometries in which transport is exactly calculable. The Levich equation gives the limiting current: i_L = 0.620·n·F·A·D^(2/3)·ν^(−1/6)·ω^(1/2)·c. The practical range for laminar flow is roughly 100–10 000 rpm; disk eccentricity and vibration corrupt the results.

Koutecký–Levich analysis

Under mixed control 1/i = 1/i_k + 1/(B·ω^(1/2)) is written. The plot of 1/i vs ω^(−1/2) of currents measured at different rotation rates is linear; the intercept gives the kinetic current (i_k = nFAk·c), the slope n·D^(2/3). In the oxygen reduction reaction (ORR) this analysis determines whether n is close to 4 (water) or 2 (peroxide).

Ring-disk electrode (RRDE)

A concentric ring is added around the disk and the two electrodes are held at independent potentials with a bipotentiostat. The intermediate generated at the disk (e.g. H₂O₂) is carried by the flow to the ring and detected there; the collection efficiency N is set by the geometry (typically 0.2–0.4) and calibrated with a standard couple such as ferricyanide. Peroxide yield and electron number: n = 4·i_D/(i_D + i_R/N). RRDE is one of the few methods sufficient on its own for measuring unstable intermediates and reaction selectivity.

Protocol in catalyst measurement

A thin-film catalyst ink is drop-cast on the glassy-carbon disk; activity is reported, after iR correction and subtraction of the background (N₂-saturated), by normalising the kinetic current at a given potential to mass (A/mg) or ECSA (A/cm²). Half-wave potential and Tafel slope are the comparison criteria.

2 Application in measurement

In RDE, controlled convection creates a constant diffusion layer; the limiting current is linear with ω^(1/2) (Levich). The Koutecký-Levich plot (1/i vs ω^(-1/2)) gives the kinetic current and the electron number. In RRDE the ring electrode collects the intermediates generated at the disk (for example H₂O₂ in ORR); a bipotentiostat controls the two electrodes independently.

Set-up: Bio-Logic BluRev or Entek RDE rotator + EC-Lab® RDE control kit or external voltage-speed control; the disk material (GC, Pt, Au) and polishing quality determine reproducibility.

For ECSA (electrochemically active surface area) and catalyst benchmarking the protocols of Garsany et al. are taken as reference.

Key equations

  • Koutecký-Levich: 1/i = 1/i_k + 1/(B·ω^(1/2))

3 Applicability

Where it applies

  • Activity, selectivity and mechanism for fuel-cell and electrolyser catalysts (ORR, HER, OER, CO₂RR).
  • Determination of diffusion coefficient and electron number; transport studies of corrosion inhibitors.

Limits and caveats

  • In gas-evolving reactions (OER, HER at high current) bubbles disturb the flow; RDE data must be interpreted with care.
  • Thick and porous catalyst layers violate the thin-film assumption; an in-layer diffusion correction is needed.

4 What did the application notes find?

  • Battery AN7 — Shows step by step the set-up of an RRDE experiment with a two-channel/bipotentiostat configuration, the simultaneous control of ring and disk channels and the determination of the collection efficiency with ferricyanide.
  • Kinetics AN56 — Shows that the Levich and Koutecký–Levich tools in the software automatically calculate n, D and k from linear sweeps taken at different rpm, and that the linearity of the plots is the criterion of method validity.
  • Electrochemistry AN66 — Shows that the diffusion coefficient can be obtained from a single rotation rate by fitting EIS at a rotating electrode with the finite-diffusion element; the result agrees with Levich analysis.
  • Fuel Cell AN11 — Explains how the electrochemically active surface area (ECSA) is calculated from the hydrogen adsorption/desorption charge on a platinum surface (210 µC/cm²); the suitable potential window and scan rate of the CV.

5 Which instruments perform this measurement?

References

  1. book A. J. Bard, L. R. Faulkner, H. S. White. Electrochemical Methods: Fundamentals and Applications (3rd ed.). Wiley, 2022.
  2. article Y. Garsany, O. A. Baturina, K. E. Swider-Lyons, S. S. Kocha. Experimental Methods for Quantifying the Activity of Platinum Electrocatalysts for the Oxygen Reduction Reaction. Anal. Chem. 82, 6321–6328, 2010. doi:10.1021/ac100306c
  3. article N. Elgrishi, K. J. Rountree, B. D. McCarthy, E. S. Rountree, T. T. Eisenhart, J. L. Dempsey. A Practical Beginner's Guide to Cyclic Voltammetry. J. Chem. Educ. 95, 197–206, 2018. doi:10.1021/acs.jchemed.7b00361 open access