Bio-LogicBio-Logic Science Instruments Authorized Distributor for Türkiye
+90 (312) 437 40 77  ·  [email protected]
TR/EN
Get a Quote
Popular topics · Getting started with EIS

PEIS, GEIS or GEIS-AA? Which control mode should be chosen and when?

Potential-controlled EIS makes it easy to define a small amplitude; low-impedance and time-variant systems, however, need current control. GEIS-AA (adaptive amplitude) combines the advantages of both.

Key points

  • Without prior knowledge, a potential amplitude of 5 - 20 mV is considered small; this is why PEIS is generally used.
  • Low-impedance systems (e.g. a Li-ion cell with a high-frequency impedance of a few mΩ): a small potential change produces a large current and the regulation loop struggles. Under current control a large current amplitude gives a small voltage response and the measurement is stable.
  • Time-variant systems: as the corrosion potential drifts, PEIS stays at the initial potential and the measurement ends up at an anodic point. GEIS (I = 0) keeps the measurement at the actual corrosion potential at all times.
  • In-situ EIS during charge or discharge can only be performed under galvanostatic control around a non-zero DC current.
  • GEIS-AA: current is controlled while the amplitude is set in volts; the instrument adapts the current amplitude to meet the target voltage amplitude. The result is the same as a potential-controlled measurement, while keeping the current control required for time-variant systems.
  • In the battery example the mean DC current during PEIS reached 166 mA but stayed at about 2 mA with GEIS: controlling zero current stresses the cell far less than controlling the initial OCP.

Figures

As the corrosion potential drifts: (a) under potential control (PC) the measurement ends up at an anodic point, (b) under current control (GC, I = 0) it is always made at the actual corrosion potential.
Figure 1. As the corrosion potential drifts: (a) under potential control (PC) the measurement ends up at an anodic point, (b) under current control (GC, I = 0) it is always made at the actual corrosion potential.
In-situ EIS during discharge: successive spectra are taken under galvanostatic control around a DC current I₁.
Figure 2. In-situ EIS during discharge: successive spectra are taken under galvanostatic control around a DC current I₁.
On a low-impedance battery, PEIS with 10 mV amplitude and GEIS with a 10 mV target give the same Nyquist spectrum.
Figure 3. On a low-impedance battery, PEIS with 10 mV amplitude and GEIS with a 10 mV target give the same Nyquist spectrum.
Mean DC current during the measurement: up to 166 mA with PEIS but about 2 mA with GEIS; current control stresses the cell less.
Figure 4. Mean DC current during the measurement: up to 166 mA with PEIS but about 2 mA with GEIS; current control stresses the cell less.

Images: Bio-Logic Science Instruments; taken from the source article.

Related application notes

Related curriculum module: Electrochemical Impedance Spectroscopy (EIS)

Source: Bio-Logic Learning Center, “PEIS or GEIS or GEIS-AA? That is the question.” · Source updated: November 19, 2024
Copyright of the images and the original text belongs to Bio-Logic SAS.

All popular topics

Would you like to set up this measurement in your laboratory?

Contact our technical team for the suitable instrument, cell and software configuration.

Quotation and technical consulting